Clamp On Ultrasonic Flow Meter Flowmasonic WUF 620 CF Series - Flow Meter Indonesia - Flow meter air, flowmeter solar, gas, Ultrasonic, Vortex, Electromagnetic

Breaking

Jumat, 24 Desember 2021

Clamp On Ultrasonic Flow Meter Flowmasonic WUF 620 CF Series

Clamp On  Ultrasonic Flow Meter Flowmasonic WUF 620 CF Series



Flowma Flowmasonic WUF 620 CF Series clamp-on transit-time ultrasonic flow meter utilizes transit-time principle. The Flowmasonic transducer is mounted outside surface of the pipe without requirements of flow stop or pipe cutting. It’s very simple, convenient for installation, calibration and maintenance.

Different sizes of transducers satisfy different measuring demand. Plus, select thermal energy measuring function to achieve completely energy analysis. It is widely applied in processing monitoring, water balance test, district heating balance test, energy efficiency monitoring as easy installation and simple operation advantages.

Flowmasonic  WUF 620 CF Series Features

  • 4 Lines Display Velocity, Flow Rate, Volume And Meter Status
  • Clamp-on Mounted, Unnecessary Pipe Cutting Or Processing Interruption
  • Acceptable Fluid Temperature Range -40°C~260°C
  • Built-in Data Storage Is Optional
  • Selecting Temperature Sensor PT1000 To Achieve Thermal Energy Measurement Function
  • Suitable For DN20-DN6000 Flow Measurement By  Selecting Different Size Transducers Bi-directional Measurement, Flow Range From 0.01m/s To 12m/s

Flowma WUF 620 CF Series 
Flowma WUF 620 CF Series

When selecting a measurement site, it is important to select an area where the fluid flow profile is fully developed to guarantee a highly accurate measurement. Use the following guidelines to select a proper installation site:

  • Choose a section of pipe that is always full of liquid, such as a vertical pipe with flow in the upward direction or a full horizontal pipe.
  • Ensure enough straight pipe length at least equal to the figure shown below for the upstream and downstream transducers installation.
  • Ensure that the pipe surface temperature at the measuring point is within the transducer temperature limits
  • .Consider the inside condition of the pipe carefully. If possible, select a section of pipe where the inside is free of excessive corrosion or scaling.

Performance Index

Measuring PrincipleTransit-time
Velocity0.01 - 12 m/s, Bi-directional Measurement
Resolution 0.25mm/s
Repeatability 0.1%
Accuracy±1.0% of Reading
Response Time0.5s
Sensitivity0.003m/s
Damping0-99s(settable by user )
Suitable FluidSuitable Fluid Clean or tiny amounts of solids, air bubbles liquid , Turbidity<10000 ppm
Power SupplyAC: 85-265V DC:12- 36V/500mA
InstallationWall Mounted
Protection ClassIP66
Operating Temperatureto +75℃
Measuring Unitmeter,ft,m³,liter,ft³,gallon,barrel etc.
Communication Output4~20mA, OCT, Relay, RS485 (Modbus-RUT), Data Logger ,GPRS
Energy UnitUnit : GJ,Opt: KWh
SecurityKeypad Lockout, System Lockout
Size244*196*114mm
WB4:C22eight2.4kg

Pages